Description
Specifications∶
Parameters: pH, Nitrite, Specific Gravity, Ascorbic Acid, Blood, Glucose, Bilirubin, Urobilinogen, Ketone, Leucocytes, Protein,Calcinm, Creatinine, Microalbumin
Measurement Principle: Dual wavelength reflectance method
Measuring Wavelength: 3 LED Wavelength (550,620,720nm)
Sample supply: manual dipping
Reaction time: 60 seconds
Throughput: Normal mode, 60 samples/hour Continuous mode, 125 samples /hour
Memory capacity: 5000 results
Strip: URIT 10G /11G urine reagent strips
Interface: Serial port(RS232), parallel port (For external printer, PS/2(For barcode reader)
Work condition: Environmental Temperature: 15-30℃, RH: 20%~80%
Power supply: Adopter 12V DC, 4.0A
Dimension: 290mm x 200mm x 100mm
Weight: 2kg
Features∶
Cold light source facility with long life and high stability
Internal thermal printer
Easy operation, cost-efficiency and high quality
Rs232 serial interface to host PC/LIS, Barcode reader optional
URIT UQ-10 & UQ-11 Urinalysis Control